Model ICHeco / ICH with humidity control.
Model ICHeco L / ICH L with humidity control and light.
Model ICH C with humidity and CO2 control.Compressor-cooled stability test chambers stand out due to their unparalleled temperature and humidity homogeneity for longterm stable ambient conditions. The climate chamber ICH has been designed for testing pharmaceutical products acc. to ICH, Q1A and Q1B, option 2 and similar global standards for stability tests of cosmetics and food.
No icing, drying out of samples or dehumidification of the chamber. Cooling aggregate and heating are situated outside the chamber in the air jacket surrounding the entire chamber thus ensuring quick and precise temperature control. The motor-driven forced air circulation, adjustable in 10 % steps, ensures homogenous temperature distribution. For supporting IQ- / OQ- / PQ validation, control can be adjusted for 3 adjustable temperature values, 2 humidity values 20 and 80 % and 3
CO2 values 5, 10 and 15 % (for ICH C) directly on the appliance.
For tests acc. to ICH Q1B, option 2, an illumination unit with a light intensity of 6000 Lux is available for model ICH L. Light sources: fluorescent lights with cold-white light (daylight: light colour 865, 6,500 K) as well as UV radiation in the spectral range of 320 to 400 nm.
Illumination unit (only model ICHeco L) acc. ICH Q1B, option 2; separately switchable via controller, one box; number of fluorescent lights with cold white light (size 110: 3, size 260 / 750: 4), light colour 865 6.500 K; number of fluorescent lights with UV lamps (all sizes: 2), spectral range from 320 to 400 nm; (daylight and UV light comply with standard illuminant D65).
AtmoCONTROL - control and logging softwareFor direct setting parameters as well as process time. Fast and easy programming of complex processes by means of the mouse or touchpad on the notebook.
- Reading out, managing and organising the data logger.
- Saving log memory in various formats.
- Online monitoring of up to 32 connected appliances.
- Optical alarms when individual alarm limits at the ControlCOCKPIT are exceeded.
- Automatic alarm to one or more email addresses.