Libra S50PC / S60PC / S70PC / S80PC:Supplied with Resolution PC software incl. quick read, quick scan, fixed wavelength, wavelength scan, quantitative analysis, kinetics, method developer and validation modules plus USB connection cable.
Libra S50 / S60 / S70 / S80:Built in software modules incl. fixed wavelength, wavelength scanning, calibration curves, kinetics, equation editor and pre-stored life science applications. Software is available in multiple languages and includes onboard data / method storage and user log on. Supplied with print via computer ("PVC") software and USB cable allowing automatic data export and printing via PC.
Wavelength range: | 190 to 1100 nm |
Wavelength reproducibility: | ± 0.1 nm |
Photometric accuracy: | ± 0.002 A at 0.5 A ± 0.004 A at 1 A ± 0.006 A at 2 A |
Photometric reproducibility: | ± 0.002 A at 1A |
Scan speed: | > 2400 nm / min |
Noise: | < 0.002 A near 2 A at 546 nm peak to peak < 0.00005 A RMS at 0 A at 700 nm and 500 nm over 20 measurements |
Libra S50PC:PC controlled scanning UV / VIS spectrophotometer.
Libra S50:Stand alone scanning UV / VIS spectrophotometer with colour touchscreen and USB interface for data / method storage.