Meter for direct analysis of moisture content acc. to the loss-on-drying method. Heating of samples with microwave radiation. Suitable for liquid and paste-like goods with moisture contents from approx. 8 to 100 %. Operation via colour display with touchscreen and 5 fixed control keys. Integrated thermal printer with permanent protocol print.
- Microwave generator: 1000 W input power, power adjustment from 2 to 100 %, selectable in 1 % steps.
- Automatic end-point determination due to mass- and moisture sensors, user-defined as loss of weight / unit of time: 1 to 50 mg / 1 to 99 s, 0.1 to 9.9 % / 1 to 99 s, 0.1 to 99.9 min.
- 320 monitoring programs, saved with mains failure protection.
- Configurable GLP protocol.
- User guidance: menu-driven, alpha-numeric dialogue text.
- Data interface: 1 x RS232 for PC, 1 x ethernet connection.
Items supplied:LMA200PM-000EU, mains cable, flat forceps, 10 disposable pipettes, 80 fibre glass fleeces, calibration weight 50 g, manual.